OM-550 u45u
APPENDIX A: CALIBRATION
•• MANUAL RESET
A Manual Reset returns the OM-550 to a "known" condition. This is useful when
troubleshooting or learning to program the unit.
A Manual Reset programs the OM-550 as follows:
• CH1 is a Voltage channel, m=1.00000, b=000000, dp=4, units = (63) "V".
• CH2 is a Type K thermocouple Temperature channel.
• CH3 through 5 are Skipped
• CH6 is Skipped but defined as CH4+CH5, decimal fix=1, units = .00 (Blank)
• The Temperature Units for all temperature channels are °C.
• The Real-time Clock/Calendar is initialized to 00:00:00 01/01/00.
• The Log Interval is 00:00:30 (30 seconds).
• The OM-550 is in the Normal (not the Calibration) mode.
• Tadjust = 00.0
• The Rejection Frequency = 60Hz.
• The data cache contains 100 "phony" points of data for channels 1 & 2.
• The OM-550 is programmed to overwrite a full cache.
•• PERFORMING A MANUAL RESET
1. Turn the unit “OFF”
2. Press & hold the 5 key
3. Momentarily press the ON key
4. Release the 5 key
A collection of messages similar to the one shown in Figure A-1will be printed.
Figure A-1